Monitoring Under Gas Anesthesia and Fixing Common Problems

Comen NM Vet Series Modular Veterinary Patient Monitor – veterinary equipment Saudi Arabia | Raed Alsharq

Good monitoring means one trained person whose main job is the patient, checking depth, oxygenation, ventilation, circulation and temperature, and writing values down at least every 5 minutes. Machines help, but they do not replace hands, eyes and ears.

Judging depth in dogs and cats

SignToo lightSurgical planeToo deep
Eye positionCentralRotated ventromediallyCentral again, pupils dilated
Palpebral reflexBriskSlow or absentAbsent
Jaw toneFirmLooseCompletely flaccid
Movement / swallowingPresentAbsentAbsent
Heart rate and BPRising with stimulationStableFalling, bradycardia, hypotension
BreathingFast, irregularRegularSlow, shallow, apneic

No single sign is enough. A central eye can mean too light or too deep: check jaw tone, reflexes and vital signs together.

Target values

ParameterTarget in dogs and catsAct when
SpO295% or above on oxygenBelow 95%
EtCO235–45 mmHgAbove 55–60 or below 30 mmHg
Mean arterial pressure60–70 mmHg or aboveMAP below 60 mmHg (Doppler systolic below 90 mmHg)
Heart rate (dog)About 60–140 /min, by sizeSudden change of 20% or more, or arrhythmia
Heart rate (cat)About 120–200 /minBelow about 100 or above about 220 /min
Temperature37–39 °CBelow 36.5 °C (start active warming earlier)
Fluid rateDogs 5 mL/kg/h, cats 3–5 mL/kg/hReassess if the total exceeds 20 mL/kg

Fluid rates follow the 2024 AAHA fluid therapy guidelines. In the CEPSAF study, pulse and pulse-oximetry monitoring were linked to lower odds of anesthetic death in cats.

Reading the capnograph

  • Normal: square waveform, baseline at 0, plateau at 35–45 mmHg.
  • Baseline above 0: rebreathing CO2 – exhausted absorbent, stuck valve, or too little flow on a non-rebreathing circuit.
  • High plateau: hypoventilation – lighten depth or ventilate.
  • Sudden drop to near 0: disconnection, tube displaced or blocked, or cardiac arrest. Check the patient immediately.
  • Sloping upstroke (“shark fin”): airway obstruction, bronchospasm or a kinked tube.

Troubleshooting table

ProblemCheck firstCommon fixes
HypotensionDepth, HR, blood loss, vasodilating drugsReduce vaporizer and add analgesia; fluid bolus if hypovolemic; treat bradycardia; vasopressor or inotrope per protocol
Low SpO2Probe position, tube placement, oxygen supply, disconnectionConfirm oxygen flow, check tube, ventilate, reposition; consider atelectasis and recruit gently
High EtCO2Depth, breathing pattern, absorbent, flowsLighten, ventilate, change absorbent, raise non-rebreathing flow
BradycardiaDepth, opioids or alpha-2s, vagal stimulation, hypothermiaReduce depth; anticholinergic if BP is also low; stop surgical traction
TachycardiaLight plane, pain, hypovolemia, hypercapniaAdd analgesia, deepen if light, assess volume and CO2
HypothermiaDuration, open body cavities, cold fluidsForced-air or circulating-water warming, warm fluids, cover extremities
Patient will not stay asleepEmpty vaporizer, wrong setting, circuit or cuff leakRefill, check dial, leak test, inflate cuff correctly, check flows
Slow recoveryTemperature, hypotension, hypoglycemia, overdoseWarm, check BP and glucose, reverse reversible drugs

Regurgitation

If you see fluid at the mouth or in the oropharynx, keep the cuff inflated, lower the head, suction the pharynx and esophagus, and lavage with water if the fluid is acidic. Keep the tube in until the patient can protect its airway.

Equipment note from Raed Alsharq: Comen veterinary monitors range from the compact ND Vet series and the pre-configured N Vet series to the modular NM Vet series, with options for capnography, invasive blood pressure and Masimo or Nellcor SpO2. We help clinics match the monitor to their caseload. Contact us.

This article is for veterinary professionals and is educational only.